EL Display |
8.3”(21.1 cm)W x 6.2”(15.8 cm)H; 160° viewing angle; Rotates 330°;Tilts 180°; Detachable; Laptop-like closure for storage and protection; Remote monitor mount (optional) |
Preferences Menu |
User may select display sweep speed (25 or 50 mm/sec), brightness (low, med., high); balloon waveform (on/off); ECG inflation markers (on/off); AP waveform auto-scaling (on/off); flashing alarms (on/off) |
ECG ECG Trigger |
Threshold dynamically adjusted by system for improved sensitivity and selectivity of the R-wave detection; Minimum = 120µV ± 20µV at normal gain; 40µV at max. gain. |
Pressure Trigger |
In automatic mode: adjusted to 50% of the difference between peak systolic and end diastolic (avg. over multiple cycles); In variable mode: User adjustable between 7 and 30 mmHg ± 3 mmHg. |
Pacer A Trigger |
R-wave detection (as above) except blanking is extended to 100 ms |
Pacer V/A-V Trigger |
V Pacer: fixed at rate up to 185 bpm (no demand pacing) A-V Pacer: fixed at ate up to 125 bpm (no demand pacing) with A-V intervals between 80-224 ms. |
Internal Trigger |
Variable mode: 40-120 bpm; Normal mode: 80 ± 1 bpm |
Tall T-Wave Rejection |
(ECG and Pacer A mode) Rejects all T-Waves where Q-T interval is <300 ms and the amplitude is <70% of QRS input amplitude |
Pacer Rejection |
(ECG and Pacer A mode) Rejects all pulses of amplitude ± 2.0 mV to ± 700 mV and durations between 0.1 ms to 2.0 ms with:1) No tail; 2) 100 ms time constant tail < 1 mV; 3) 25 ms time constant tail < 1 mV; 4) 4 ms time constant tail < 2 mV |
ECG Leads |
I, II, III, AVR, AVL, AVF, V (12 lead compatibility) On-keypad switch between direct leads and external monitor |
ECG Gain (default) |
1 V output per 1 mV input ±5% (waveform automatically scaled to occupy ECG display window) |
Gain (variable) |
0.15 cm to 3.0 cm/mV ±20% (autoscaling disabled) |
Frequency Response |
0.5-12 Hz (display); 0.5-135 Hz (Output to Monitor) |
Defibrillator Protection |
Discharge levels =360 J (trace returns to screen in 2 sec. max) |
ESIS |
Automatic suppression with internal ECG amplifier |
AP Source Selection |
On-keypad switch (transducer/external) |
|
|
AP Input Sensitivity |
5.0µV/V/mm Hg |
Excitation |
+5 VDC ± 5% @ 130 mA (max) |
Frequency Response |
0.5 to 12 Hz (-3 dB max atten.) |
Timing Modes |
Automatic (R-Trac ON or OFF); Manual |
ECG Connector |
6-pin 5-lead AAMI standard |
Pressure Connector |
6-pin AAMI standard |
External Mon. Inputs: |
ECG: 1 V/mV (nominal); AP: 1 V/100 mm Hg (nominal) |
ECG/BP Output |
Phone Jack: ECG 1V / 1 mV, BP 1V/100 mm Hg |
Mains Voltage/Freq |
100-120 VAC ± 10% or 220-240 VAC ± 10% ; 50/60 Hz ± 3 Hz |
Internal Battery |
24 VDC (nominal), 17.2 Amp-hour, 2.25 hrs. (min. @ 120 bpm) |
Type |
Maintenance free; Sealed lead-acid |
System Compressor |
Dual head diaphragm pump with brushless DC motor |
IAB Helium (He+) |
Medical-grade; 0.5 L (2900 psi) or 0.69 L (2200 psi) |
He+ Endurance (nom.) |
1.8 mo. (.5 L); 2 mo. (0.69 L): continuous operation - 24 hrs./day |
Condensate Removal |
Fully automatic condensate removal and disposal |
PC-IABP (optional) |
Software for remote clinical assistance and training |
Service Diagnostics |
Built-in software for system analysis and troubleshooting |
Modem Data Rate |
Up to 9,600 baud |
Modem Certifications |
Registered with FCC and accepted in =89 countries |
Size on Cart |
43.3”H x 16.8”W x 22.3”D (109 cm x 42.7 cm x 56.6 cm) |
Size off Cart |
26.9”H x 10.8”W x 20.5”D (68.3 cm x 27.4 cm x 52.1 cm) |
Console Weight |
76.8 lbs.(34.8 Kg) |
Monitor |
9.45 lbs. (4.3 Kg) |
Hospital Cart Weight |
61.0 lbs. (27.7 Kg) |
Internal Battery |
34 lbs. (15.4 Kg) |
UTS Version |
Rugged base which attaches to docking station (DS) |
Lightweight DS |
7.8”(45.1 cm) x 20.0”(50.8 cm) (Optional) |
Basic DS |
21.5”(54.6 cm) x 21.5”(54.6 cm) |
Ext Battery |
24 VDC (nominal), 17.2 Amp- hour, 1.8 hrs. (min. @ 120 bpm) (Optional) |
Aux. DC Input |
20.5-35.0 VDC |
Shock and Vibration |
ECRI AIII.3.1, 3.2.2, 3.3, 3.4 Mil Std 810E method 514.4, Cat. I; RTCA DO-160C, 1989, Section B, Curve N |
Op. Temp. |
10°C - 40°C |
Op. Humidity |
5 - 95% (R.H.) non-condensing |
Op. Altitude |
Up to 12,000 feet (3,657 m); automatic altitude correction for IAB pressure |
Printer Type |
Thermal array - 50 mm wide |
Printer Menu |
On-screen selection of waveforms; strip length; timed print; print on alarm; trend data; and alarm/trigger log |
Waveforms |
(any 1 or 2) ECG (lead number & size labeled), arterial pressure, BPW |
Event Markers |
Trigger pulse (upper), inflation interval |
Trend Graphs |
<=8 hrs of heart rate; peak systolic, end diastolic, mean, augmented diastolic |
Additional Data |
Space for patient I.D.; Date; Time; Alarm and Advisory messages; Timing mode; Trigger; HR, Frequency; Syst/Dias (in 1:1); Assisted and Unassisted
Syst/Dias (in 1:2 or 1:3); Augmented Pressure; Mean |
Doppler |
8 MHZ non-directional probe |
Saline Pole |
Height adjustable |
Simulator* |
Emulates ventricular rhythms; Pacers; HR; A-fib |
Remote Mount* |
Mounts keypad/monitor on bypass pump |
Storage Case* |
3-pocket design holds cables and spare items |
Pump Cover* |
Padded vinyl with pocket |
|
|
*(optional) |
|